Keysight / Agilent E1433A / E1433B 8-Channel 196 kSa/s Digitizer Plus DSP
  • Agilent_E1433B_View1

Keysight / Agilent E1433A / E1433B 8-Channel 196 kSa/s Digitizer Plus DSP

Features

  • 1-Slot, C-size VXI Module
  • 8-Channel 196 kSa/s Digitizer Plus DSP
  • Transducer Signal Conditioning
  • Anti-Alias Protection
  • High Speed Measurement Computation
  • Standard Configuration has 4 MB RAM

View our selection of these items below.


Have a quote from somewhere else? Click Here




Contact Us
Share:



Opt ANC: 32MB RAM
Opt AYF: Tachometer Input
Opt UGV: Local bus interface installed






Opt 1D1: Povides real-time 1/1 or one-third octave measurements. Third octave bands cover 3.15 Hz to 20 kHz and the Full Octave bands cover 4 Hz to 16 kHz. All eight channels of the E1433B can be performing real-time octave measurements simultaneously
Opt ANC: 32MB RAM
Opt AYF: Tachometer Input
Opt UGV: Local bus interface installed






Opt 1D1: Povides real-time 1/1 or one-third octave measurements. Third octave bands cover 3.15 Hz to 20 kHz and the Full Octave bands cover 4 Hz to 16 kHz. All eight channels of the E1433B can be performing real-time octave measurements simultaneously
Opt 1D4: Additional Arbitrary Source installed
Opt ANC: 32MB RAM
Opt UGV: Local bus interface installed






This Agilent / HP E1433B 8-Channel 196 kSa/s Digitizer Plus DSP is used and in excellent condition.

Installed Options: None (standard configuration)






Opt 1DL: One Input Assembly (4-Channels)
Opt UGV:
Local Bus Interface Installed






Opt 1D1: Povides real-time 1/1 or one-third octave measurements. Third octave bands cover 3.15 Hz to 20 kHz and the Full Octave bands cover 4 Hz to 16 kHz. All eight channels of the E1433B can be performing real-time octave measurements simultaneously
Opt AYF: Tachometer Input
Opt UGV: Local bus interface installed






HP E1432-61600 Voltage-type Break Out Box
Each of the eight connectors in this type of Break Out Box is connected to an independent, floating current source. These are intended to power integrated-circuit piezo-electric (ICP) transducers. They supply 4.5 mA (nominal) at up to 28 volts. The current sources are controllable by software in groups of four. That is, the current sources for connectors 1-4 can be turned on or off as a group, as can the current sources for connectors 5-8.

Break Out Box Grounding
Each connector on the HP E1432-61600 and HP E1432-61602 Break Out Box has a small manual switch next to it. When this switch is in the "GND" position the outer shell of the connector is grounded to the chassis ground of the VXI mainframe. When it is in the "DIFF" position it is not grounded to the mainframe and will float if not grounded elsewhere in the system (such as at the sensor). The connector shell should not be allowed to float: if the switch is in the "DIFF" position the shell should be grounded elsewhere in the system.







HP E1432-61602 Voltage-type Break Out Box
In this type of Break Out Box the signal is sent straight through to the sub-miniature "D" connectors on the HP E1432A/33A.

Break Out Box Grounding
Each connector on the HP E1432-61600 and HP E1432-61602 Break Out Box has a small manual switch next to it. When this switch is in the "GND" position the outer shell of the connector is grounded to the chassis ground of the VXI mainframe. When it is in the "DIFF" position it is not grounded to the mainframe and will float if not grounded elsewhere in the system (such as at the sensor). The connector shell should not be allowed to float: if the switch is in the "DIFF" position the shell should be grounded elsewhere in the system.







HP E1432-61602 Voltage-type Break Out Box
In this type of Break Out Box the signal is sent straight through to the sub-miniature "D" connectors on the HP E1432A/33A.

Break Out Box Grounding
Each connector on the HP E1432-61600 and HP E1432-61602 Break Out Box has a small manual switch next to it. When this switch is in the "GND" position the outer shell of the connector is grounded to the chassis ground of the VXI mainframe. When it is in the "DIFF" position it is not grounded to the mainframe and will float if not grounded elsewhere in the system (such as at the sensor). The connector shell should not be allowed to float: if the switch is in the "DIFF" position the shell should be grounded elsewhere in the system.







Opt 1D4: Additional Arbitrary Source installed
Opt 1DL: One Input Assembly (4-Channels)






Opt 1D1: Povides real-time 1/1 or one-third octave measurements. Third octave bands cover 3.15 Hz to 20 kHz and the Full Octave bands cover 4 Hz to 16 kHz. All eight channels of the E1433B can be performing real-time octave measurements simultaneously
Opt 1D4: Additional Arbitrary Source installed
Opt UGV: Local bus interface installed





  • Description
  • Tech Specs
  • Datasheets / Manuals 
Contact Us
Share:

Description

The Agilent E1433A & E1433B 8-Channel 196 kSa/s Digitizer plus DSP are C-size VXI modules. "196 kSa/s" refers to the maximum sample rate of 196,608 samples per second per channel. The E1433A & E1433B may contain either one or two four-channel input assemblies so that the modules may have a total of up to eight inputs. This modules integrate transducer signal conditioning, anti-alias protection, digitization and high speed measurement computation in a single-slot VXI card. Onboard digital signal processing and 32 Mbytes of RAM maximizes total system performance and flexibility.

Product Family: E1433A, E1433B
Contact Us
Share:

Specifications