Features:
- Alarm for Low Water, Temperature and Pump/Motor Overloading
- Excessive Temperature Protection Dial
- Fault Displays for Alarms
- RS232 and RS485 Remote Interfaces
All units fulfill the requirements of safety class 2 according to DIN 12879 and are thus suitable for unsupervised continuous operation.
The circulator pump motor is protected against thermal overloading. All temperature sensors are permanently monitored according to break or short circuit. The cooling machine is integrated into the general safety circuit. The control of the pre-selected temperature is carried out automatically via the Thermo Haake FuzzyStar control.
The comprehensive safety system is designed on the principle of the concept of the "single fault" (EN 61 01 0). This assumes that two separate faults do not occur simultaneously. This system therefore offers protection against one (single) fault. Safety class 2 according to DIN 12879 A variably adjustable excess temperature protection and independent low liquid level protection which is preset to the lowest level allows the usage of different heat transfer liquids.
F8 Specifications:
• Working Temperature Range: 35 - 250°C
• With Tap Water Cooling: 20 - 250°C
• With Other Cooling: -60 - 250°C
• Temperature Accuracy: 0.01±K
N8 Specifications:
• Working Temperature Range: 45 - 300°C
• With Tap Water Cooling: 20 - 300°C
• With Other Cooling: -60 - 300°C
• Temperature Accuracy: 0.01±K
The F8 Thermo Circulator may still be available for purchase and support from Artisan Technology Group beyond End-Of-Life (EOL) by the manufacturer (OEM).
Artisan Scientific Corporation dba Artisan Technology Group is not affiliated with or a distributor for Thermo. The depiction, description, or sale of products featuring names, trademarks, brands, and logos is for identification and/or reference purposes only and does not indicate any affiliation with or authorization by any rights holder.