This unit is designed to work with an external vacuum pump. The thermal unit does not lower, but it does raise and pump. Before installing and operating this unit, the coolant lines will need to be completely flushed. In order for unit to operate correctly through the full temperature range, 3M Novec HFE7500 Engineered Fluid needs to be installed. This unit comes with a quick-connect air hose. Also, the unit will come in a custom wooden crate.
The environmental chamber has an application specific thermal unit installed. The thermal unit or tool is the part that makes contact with the Device Under Test, and applies the change in temperature. The tool may need to be changed according to the specific test being performed.
View Photo of the tool installed in the environmental chamber.
Configuration:
• Input: 190-230 VAC
• Frequency: 50-60 Hz
• Phase: 3
• Interrupt Current: 65,000 AIC
• Largest Load Ampere Rating: 33 A
• Machine Main Breaker Ampere Rating: 50 A
• Full Load Amps: 40 A
• Temperature Range (Passive): -30°C to +150°C
• Temperature Ramp Rate: 100°C Per Second
• ESD: <100 V within 6" of device
• Communication Interface: GPIB and RS232
• Air: 80 psi, 2-3 scfm, clean, dry dewpoint 15°C below coldest chiller setpoint
• Weight: 340 kgs (750 lbs)
• Dimensions: 25.4W x 42D x 63H inches (0.65m x 1.08m x 1.62m)
• Chiller: Affinity PAG-040K-BE51CBD2 Installed*
• Passive Model Version
Computer:
• 700 MHz Intel PIII Processor
• 256 MB RAM
• 18 GB HDD
• ETC-2000 Software v2.2.1
*If desired, the Affinity PAG-040K-BE51CBD2 Chiller can be sold individually. Please contact us for pricing and availability.