Delta Design ETC 2000L High Performance Logic IC Thermal Characterization Unit

Stock # 73574-4
The thermal unit does not raise and lower. Before installing and operating this unit, the coolant lines will need to be completely flushed. In order for unit to operate correctly through the full temperature range, 3M Novec HFE7500 Engineered Fluid needs to be installed. This unit comes with a quick-connect air hose.

The environmental chamber does not have the application specific thermal unit installed. The thermal unit or Tool is the part that makes contact with the Device Under Test, and applies the changes in temperature. A tool will need to be installed according to the specific test being performed. View Photo of the tool installed in the environmental chamber.

Configuration:
• Input: 190-230 VAC
• Frequency: 50-60 Hz
• Phase: 3
• Interrupt Current: 65,000 AIC
• Largest Load Ampere Rating: 33 A
• Machine Main Breaker Ampere Rating: 50 A
• Full Load Amps: 40 A
• Temperature Range (Passive): -30°C to +150°C
• Temperature Ramp Rate: 100°C Per Second
• ESD: <100 V within 6" of device
• Communication Interface: GPIB and RS232
• Air: 80 psi, 2-3 scfm, clean, dry dewpoint 15°C below coldest chiller setpoint
• Weight: 340 kgs (750 lbs)
• Dimensions: 25.4W x 42D x 63H inches (0.65m x 1.08m x 1.62m)
• Chiller: Affinity PAG-040K-BE51CBD2 Installed*
Optional Delta Design Power Summing Module Installed
• Passive Model Version

Computer:
• 700 MHz Intel PIII Processor
• 128 MB RAM
• 37 GB HDD
• ETC-2000 Software v2.2.1

*If desired, the Affinity PAG-040K-BE51CBD2 Chiller can be sold individually. Please contact us for pricing and availability.

View Photo of the interior with front door open.

View Photo of the interior with right hand side door open.

View Photo of the interior with left hand side door open.

View Photo of the interior with back door open.

View Photo of the back door.

Show More Description
Delta Design ETC 2000 Series High Performance Logic IC Thermal Characterization Unit
Contact Us
Fax: